Socorex® classic 187 automatic syringes: Cleaning and Drying
Автор: Socorex Isba SA
Загружено: 2020-10-19
Просмотров: 4866
This video shows the priming, disassembling, cleaning and drying of the Socorex® classic 187, fully autoclavable veterinary injection syringe.
When dosing is completed, remove feed sources and activate several times in order to eliminate any residual liquid.
Rinse, disassemble, clean and dry the instrument parts according to the instructions in the ‘‘maintenance’’ chapter.
Never store a syringe without prior purging, cleaning and drying.
How to clean the automatic syringe:
The self-refilling syringe and accessories must be carefully cleaned and dried after each use.
Aspirate an appropriate cleaning solution (water or detergent) and rinse thoroughly.
Activate the syringe several times with air to remove residual liquid before disassembling.
How to disassemble and clean the automatic syringe:
Remove needle, tubing and/or vial holder.
Hold handle and unscrew casing in order to access the glass barrel.
Push rod to free plunger.
Unscrew aspiration and dispensing valves.
Dry the valve assembly (seat, ball), syringe body, barrel and plunger before reassembling and storing.
Find more information on our website: https://www.socorex.com/en/veterinary
Here you can find the original oprating instructions:
https://www.socorex.com/images/Docume...