# 1 Most Common Problem in HPLC: Air Bubble in the Pump
Автор: Axion Labs & Training Institute
Загружено: 2022-10-18
Просмотров: 38957
Identifying Air Bubbles: Learn to recognize the symptoms of an air bubble in your HPLC pump, such as unusual pressure ripple and fluctuating UV signals.
Simple Fixing Method: We demonstrate a straightforward procedure to eliminate air bubbles, including opening the purge valve and adjusting the flow rate.
Verifying Solution Effectiveness: Understand how to check if the air bubble has been successfully removed by observing the pressure ripple and ensuring it falls within the acceptable range.
Software Tips: Gain insights into using diagnostic tools in HPLC software to monitor and confirm the pressure ripple status.
Read the blog here: https://axionlabs.com/chromatography-...
Subscribe for more insights into chromatography techniques and applications.
🔗 Follow Us:
Website: https://axionlabs.com/
Facebook: / axionlabs
LinkedIn: / axion-labs
Axion offers hands-on HPLC (High-performance liquid chromatography or high-pressure liquid chromatography) and GC (gas chromatography) training courses. Gain advanced HPLC & GC skills with our 5 day hands-on LC/GC bootcamp. If you can't make it to Chicago choose one of our online courses. After taking the LC/GC bootcamp further advance your skills by taking the advanced topics courses.