SSCOR VX-2 Portable Suction Unit: How To Replace The Disposable Suction Liner
Автор: EO Medical
Загружено: 2021-08-16
Просмотров: 855
SSCOR VX-2 is a powerful, emergency, portable suction pump device specifically designed for aspiration of airway secretions.
This video is a quick user guide on how to replace the disposable suction liner of SCCOR VX-2 Portable Aspirator, created by EO Medical in Singapore
Introduction:
The disposable suction liner including the patient tubing are intended for single patient use. Replace the suction liner and patient tubing in accordance to your hospital Infection Prevention and Control Guidelines
Steps:
Before removing the liner,
disconnect all tubings.
Twist off the detachable stopper located on the rim.
Plug the stopper into the patient port.
This prevents any accidental liquid spillage when removing the suction liner.
Using the convenient handles located on the sides of the liner, remove the liner from the canister.
Clean and disinfect the portable suction unit in accordance to your hospital Infection Prevention and Control Guidelines.
Place the new liner into the canister and push firmly.
The replacement is complete.
Dispose the used suction liner and patient tubing in accordance to your hospital Infection Prevention and Control Guidelines.
TIMESTAMPS:
0:00 Introduction
0:10 See suction power of SCCOR VX-2
0:21 Disconnecting tubings from suction liner
0:33 How to stopper the suction liner's opening
0:58 How to remove the suction liner from canister
1:19 SCCOR surface cleaning recommendations
1:25 Install a new suction liner
See other tutorial videos in the SCCOR VX-2 series here:
Full Playlist - • SSCOR VX-2 Suction Pump
How To Perform Pre-Use Checks - • SSCOR VX-2 Portable Suction Unit: How To P...
How To Replace The Disposable Suction Liner - • SSCOR VX-2 Portable Suction Unit: How To R...
This material is provided as a supplementary learning tool to the User Manual. It is not intended to replace the User Manual which contains important information on ‘How to use a Product’, including safety symbols.
Happy to assist!
Find out more: https://eo.com.sg/product/sscor-vx-2-...